Abstract
From validation studies, it was possible to estimate a measurement uncertainty of several elements such as Al, Ba, Ca, Cu, Cr, Cd, Fe, Mg, Mn, Ni and K in water samples from Guarapiranga Dam. These elements were analyzed by optical emission spectrometry with inductively coupled plasma (ICP-OES). The value of relative estimated uncertainties were between 3% and 15%. The greatest uncertainty contributions were analytical curve, and the recovery method, which were related with elements concentrations and the equipment response. Water samples analyzed were compared with CONAMA Resolution #357/2005.
Cite
CITATION STYLE
Faustino, M. G., Marques, J. R., Monteiro, L. R., Stellato, T. B., Soares, S. M. V., Silva, T. B. S. C., … Cotrim, M. E. B. (2016). Uncertainty estimation in the determination of metals in superficial water by ICP-OES. In Journal of Physics: Conference Series (Vol. 733). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/733/1/012032