Electrochemical biosensor for simplified determination of salivary uric acid

19Citations
Citations of this article
19Readers
Mendeley users who have this article in their library.

Abstract

A uric acid biosensor aimed at a simplified determination of salivary uric acid was fabricated and utilized for the measurement of the diurnal variation of salivary uric acid. The biosensor measures uric acid as the change in the amount of hydrogen peroxide produced in the uricase reaction. Because uric acid is oxidized as easily as hydrogen peroxide, the osmium-HRP redox reaction was employed. The sensitivity of the biosensor was 170 nA/mM, which was sufficient for salivary uric acid determination. For simplified measurement of a saliva sample, a paperbased saliva sampling device, which enables the sample collection of a regulated amount of saliva in 5 s, was used. As a test using an actual sample, the diurnal variation of salivary uric acid was measured. The result indicated that salivary lactic acid increased in the morning. The total measurement time for the saliva measurement was approximately 3 min, which was sufficiently fast for the purpose of daily health management. The proposed method is expected to be used not only in gout treatment but also possibly in the measurement of other substances contained in saliva.

Cite

CITATION STYLE

APA

Kudo, H., & Takagi, Y. (2018). Electrochemical biosensor for simplified determination of salivary uric acid. Sensors and Materials, 30(5), 1187–1195. https://doi.org/10.18494/SAM.2018.1886