Introduction: The cementless Oxford partial knee arthroplasty is associated with low perioperative complications and good long-term survival rates. However, perioperative fractures remain a serious morbidity for patients. Case Report: This case report describes an early post-operative tibial periprosthetic fracture through the keel slot, which we believe may be contributed by the deep implant keel design and the presence of a narrow metaphysis in the Asian knee. The patient subsequently underwent a revision total knee replacement and fixation of the periprosthetic fracture. Conclusion: This paper discusses the ways to identify patients at high risk of developing periprosthetic fractures and to minimize such occurrences, including adopting a modified tibial preparation, doing precise saw cuts, and considering a cemented tibial implant. Keywords: Cementless, Oxford, unicompartmental, periprosthetic, keel.
CITATION STYLE
Xian, L. Z., & Chye, A. T. H. (2021). An Early Periprosthetic Fracture of a Cementless Oxford Unicompartmental Knee Arthroplasty: Risk Factors and Mitigation Strategies. JOURNAL OF ORTHOPAEDIC CASE REPORTS, 11(4). https://doi.org/10.13107/jocr.2021.v11.i04.2156