Abstract

We compared the newly approved BacT/Alert Virtuo blood culture system to the BacT/Alert 3D system using 115 clinical bacterial and fungal isolates in 784 simulated blood culture bottles. The time to detection was reduced by roughly 20% in the Virtuo system (P<0.0001) while the detection rate did not differ.

Cite

CITATION STYLE

APA

Altun, O., Almuhayawi, M., Lüthje, P., Taha, R., Ullberg, M., & Özenci, V. (2016). Controlled evaluation of the new BacT/alert virtuo blood culture system for detection and time to detection of bacteria and yeasts. Journal of Clinical Microbiology, 54(4), 1148–1151. https://doi.org/10.1128/JCM.03362-15