Abstract
Background: Hospitals in tropical countries experience conditions that exceed manufacturer temperature and humidity limits for point-of-care (POC) glucose reagents. Our goal was to assess the effects of out-of-limits storage temperature, operating temperature, and operating humidity on POC glucose measurement reliability. Methods: Quality control measurements were performed monthly using glucose test strips stored under controlled conditions and in inpatient wards under ambient conditions. Glucose test strips were evaluated in groups organized by operating temperatures of 24-25 (group 1), 28-29 (group 2), and 33-34°C (group 3), and relative humidity (RH) of ≤70 (group A), ∼80 (group B), and ∼90% (group C). Results: Glucose results for different storage conditions were inconsistent. Measurements at higher operating temperatures had lower values with mean differences of -2.4 (P < .001) and -36.5 (P < .001) and -37.4 (P < .001) and -13.2 (P < .001) and -16.6 (P
Author supplied keywords
Cite
CITATION STYLE
Pratumvinit, B., Charoenkoop, N., Niwattisaiwong, S., Kost, G. J., & Tientadakul, P. (2016). The effects of temperature and relative humidity on point-of-care glucose measurements in hospital practice in a tropical clinical setting. Journal of Diabetes Science and Technology, 10(5), 1094–1100. https://doi.org/10.1177/1932296816633485