Abstract
A multisyringe flow injection analysis (MSFIA) method for the determination of tinidazole (TNZ) in pharmaceutical preparations was developed. TNZ concentrations were determined by spectrophotometric detection at 374 nm wavelength using a light emitting diode (LED). The MSFIA parameters were optimized. Under the optimized experimental conditions, a linear dynamic range from 2 to 56 mg L-1 TNZ, a 3σ detection limit of 0.3 mg L -1, and a 10σ quantification limit of 1.2 mg L-1 were obtained. The relative standard deviation (%R.S.D.) at 2, 28 and 56 mg L-1 was 2.3, 1.4 and 0.9%, respectively (n = 10). A sample throughput of 76 determinations/h was achieved.
Author supplied keywords
Cite
CITATION STYLE
Guzmán-Mar, J. L., Hinojoza-Reyes, L., Hernandez-Ramirez, A., Peralta-Hernandez, J. M., López-Chuken, V. J., López-De-alba, P. L., & Cerda-Martín, V. (2010). Automatic multisyringe flow injection system for the spectrophotometric determination of tinidazole in pharmaceutical preparations. Journal of the Chilean Chemical Society, 55(2), 215–218. https://doi.org/10.4067/S0717-97072010000200015