Abstract
This study aimed to assess the cyclic fatigue resistance of Genius and EdgeFile X1 reciprocating instruments compared with WaveOne Gold Primary. Twenty Genius (Ultradent) 25.04, 20 Genius 30.04, 20 EdgeFile X1 (EdgeEndo) and 20 WaveOne Gold Primary (Dentsply Maillefer) instruments were included in this study and tested in a static cyclic fatigue testing device, which has an artificial stainless steel canal with a 60 angle of curvature and a 5-mm radius of curvature. All instruments were operated in reciprocation mode until fracture occurred. The number of cycles to failure (NCF) was calculated and time to fracture (TF) was recorded in seconds using a digital chronometer. The mean and standard deviations of NCF and TF were calculated for each reciprocating system and the data were subjected to Kruskal-Wallis one-way analysis of variance and to Dunn's test (p
Author supplied keywords
Cite
CITATION STYLE
Bueno, C. R. E., Cury, M. T. S., Vasques, A. M. V., Sivieri-Araújo, G., Jacinto, R. C., Gomes-Filho, J. E., … Dezan, E. (2019). Cyclic fatigue resistance of novel Genius and Edgefile nickel-titanium reciprocating instruments. Brazilian Oral Research, 33. https://doi.org/10.1590/1807-3107BOR-2019.VOL33.0028