Effect of Magnetic Resonance Imaging on the Motion Accuracy of Magnetic Resonance Imaging‑compatible Focused Ultrasound Robotic System

4Citations
Citations of this article
7Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Purpose: The current study provides insights into the challenges of safely operating a magnetic resonance imaging (MRI)-guided focused ultrasound (MRgFUS) robotic system in a high-field MRI scanner in terms of robotic motion accuracy. Materials and Methods: Grid sonications were carried out in phantoms and excised porcine tissue in a 3T MRI scanner using an existing MRgFUS robotic system. Fast low-angle shot-based magnetic resonance thermometry was employed for the intraprocedural monitoring of thermal distribution. Results: Strong shifting of the heated spots from the intended points was observed owing to electromagnetic interference (EMI)-induced malfunctions in system’s operation. Increasing the slice thickness of the thermometry sequence to at least 8 mm was proven an efficient method for preserving the robotic motion accuracy. Conclusions: These findings raise awareness about EMI effects on the motion accuracy of MRgFUS robotic devices and how they can be mitigated by employing suitable thermometry parameters.

Cite

CITATION STYLE

APA

Antoniou, A., Evripidou, N., Chrysanthou, A., Georgiou, L., Ioannides, C., Spanoudes, K., & Damianou, C. (2024). Effect of Magnetic Resonance Imaging on the Motion Accuracy of Magnetic Resonance Imaging‑compatible Focused Ultrasound Robotic System. Journal of Medical Physics, 49(2), 203–212. https://doi.org/10.4103/jmp.jmp_7_24

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free