Abstract
The predictors of pacing capture threshold (PCT) exacerbation after leadless pacemaker implantation re-main unknown. We analyzed the predictors of poor PCT by identifying risk factors using multivariate logistic regression analysis for 211 patients with leadless pacemaker implantation. Twenty patients met the criteria for elevated PCT levels and were categorized in the poor PCT group. Multivariate analyses revealed that PCT (P < 0.0001) and pacing impedance (P = 0.03) were independent predictors of PCT exacerbation. Elevated PCT levels and low pacing impedance during leadless pacemaker implantation were potential risk factors for the re-placement of leadless pacemakers after the procedure.
Author supplied keywords
Cite
CITATION STYLE
Morita, J., Kondo, Y., Okada, T., Kitai, T., Kasai, Y., & Fujita, T. (2023). Predictors of Pacing Capture Threshold Exacerbation After Leadless Pacemaker Implantation. International Heart Journal, 64(4), 602–605. https://doi.org/10.1536/ihj.22-698