Impact of the counts density of calibration image and the image reconstruction parameters in 131I SPECT image quantification

  • Carvalho S
  • Costa A
  • Ramos C
  • et al.
N/ACitations
Citations of this article
12Readers
Mendeley users who have this article in their library.

Abstract

Iodine-131 (131 I) has been used for diagnosis and therapy in Nuclear Medicine Centers in Brazil for more than 50 years. The present study aims to investigate the impact of the counts density and the reconstruction parameters in the calibration factor determination and in the image quantification, considering the reality of Brazilian dosimetry studies. For this task, images were quantified using calibration images with high and low counts density and reconstructed adopting two different parameters approaches, usually employed in patients images. SPECT quantification results presented in this work follow other previous 131 I SPECT studies and suggest that, due to the long time interval between the first e last images, as required by the Brazilian guideline, the image quantification accuracy can be improved if the counts density of calibration images is considered.

Cite

CITATION STYLE

APA

Carvalho, S. M., Costa, A. P. M., Ramos, C. D., Brunetto, S. Q., & Bonifácio, D. A. B. (2018). Impact of the counts density of calibration image and the image reconstruction parameters in 131I SPECT image quantification. Brazilian Journal of Radiation Sciences, 6(1). https://doi.org/10.15392/bjrs.v6i1.332

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free