Abstract
An updated liquid chromatographic (LC) method for the determination of glyphosate in technical material and formulations was tested by 7 laboratories. Samples were dissolved in water and injected into an LC system with a C18 guard column, an anion exchange analytical column, and a phosphate buffer mobile phase with 12 to 15% methanol. Detection was by UV absorption at 195 nm or refractive index. Automated injections were made. Calculations were based on peak area comparisons with external standards. The interlaboratory study analyzed 6 matched pairs of 3 glyphosate technical materials and 3 glyphosate formulations. The matched pair results were analyzed by using the AOAC spreadsheet and Youden's matched pair calculations. Coefficients of variation for the 6 samples ranged from 0.46 to 0.85%.
Cite
CITATION STYLE
Gavlick, W. K., & Tomkins, D. F. (2008). An updated liquid chromatographic assay for the determination of glyphosate in technical material and formulations. Journal of AOAC International, 91(1), 1–4. https://doi.org/10.1093/jaoac/91.1.1
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.