Abstract
A new resistive bolometer system has been developed for MAST-Upgrade. It will measure radiated power in the new Super-X divertor, with millisecond time resolution, along 16 vertical and 16 horizontal lines of sight. The system uses a Xilinx Zynq-7000 series Field-Programmable Gate Array (FPGA) in the D-TACQ ACQ2106 carrier to perform real time data acquisition and signal processing. The FPGA enables AC-synchronous detection using high performance digital filtering to achieve a high signal-to-noise ratio and will be able to output processed data in real time with millisecond latency. The system has been installed on 8 previously unused channels of the JET vertical bolometer system. Initial results suggest good agreement with data from existing vertical channels but with higher bandwidth and signal-to-noise ratio.
Cite
CITATION STYLE
Lovell, J., Naylor, G., Field, A., Drewelow, P., & Sharples, R. (2016). An FPGA-based bolometer for the MAST-U Super-X divertor. In Review of Scientific Instruments (Vol. 87). American Institute of Physics Inc. https://doi.org/10.1063/1.4961556