Detection and correction of frequency instabilities for volumetric 1H echo-planar spectroscopic imaging

64Citations
Citations of this article
67Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Spectral quality in 1H magnetic resonance spectroscopic imaging (MRSI) critically depends on the stability of the main magnetic field. For echo-planar MRSI implemented at 3 T, temperature variation in the passive steel shims of the magnet system can lead to a significant drift in the resonance frequency. A method is presented that incorporates interleaved measurement of the instantaneous resonance frequency of a reference water signal into a volumetric MRSI sequence and allows correction for the drift during postprocessing. Results from normal human brain at 3 T indicate that the correction largely removes lineshape distortions, recovers metabolite signal loss, and improves spectral quality by reducing the width of spectral lines; however, particularly in inferior regions, other sources of distortion may be present that cause broadening of spectral lines. © 2005 Wiley-Liss, Inc.