Performance evaluation of alternate ESR measurement method using BC-780 automated hematology analyzer: a comparison study with the Westergren reference method

4Citations
Citations of this article
17Readers
Mendeley users who have this article in their library.

Abstract

Objectives: Implementation of alternate erythrocyte sedimentation rate (ESR) measurement method is increasing worldwide due to its various advantages. In this study, we aim to evaluate the analytical performance of the BC-780 automated hematology analyzer in measurement of ESR value. Methods: Analyzer performance including precision study, carryover, sample stability and potential interferences are examined. Samples with ESR values spanning the whole analytical ESR range are included for method comparison study. Samples with different hematocrit (Hct) and mean corpuscular volume (MCV) values are also analyzed and compared with the results obtained from the Westergren reference method. Results: Precisions and carryover results are consistent with the manufacturers’ claim. ESR values do not change significantly in the samples stored at 2–8 °C for 24 h (h) or at room temperature (RT) for 8 h, but significantly decreased (p<0.001) when stored at RT for 24 h. Significant increase in ESR value is documented in samples that are hemolyzed (hemoglobin concentration ranged from 1.28–6.01 g/L) (p=0.010) or lipemic (triglyceride above 4.75 mmol/L) (p=0.001). Method comparison study yields a proportional difference with a regression equation=3.08+ 0.98x. Bland–Altman analysis shows a mean absolute bias of 3.12 mm. The obtained absolute mean biases are below 5 mm in all analytical categories except for the group where MCV>100 fL. Conclusions: Most tested parameters met the manufacturer’s specifications and were comparable to the reference method. Despite the presence of positive bias, it falls within acceptable criteria. Extensive validation against potential interferences such as hemolysis/lipemia is still necessary in future.

Cite

CITATION STYLE

APA

Prompetchara, E., Parnsamut, C., Wangviwat, N., Pitakpolrat, P., Chaiwong, K., Limpornpukdee, O., … Ketloy, C. (2024). Performance evaluation of alternate ESR measurement method using BC-780 automated hematology analyzer: a comparison study with the Westergren reference method. Clinical Chemistry and Laboratory Medicine, 62(2), 303–311. https://doi.org/10.1515/cclm-2023-0499

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free