Compact CH4 sensor system based on a continuous-wave, low power consumption, room temperature interband cascade laser

140Citations
Citations of this article
56Readers
Mendeley users who have this article in their library.

Abstract

A tunable diode laser absorption spectroscopy-based methane sensor, employing a dense-pattern multi-pass gas cell and a 3.3 μm, CW, DFB, room temperature interband cascade laser (ICL), is reported. The optical integration based on an advanced folded optical path design and an efficient ICL control system with appropriate electrical power management resulted in a CH4 sensor with a small footprint (32 × 20 × 17 cm3) and low-power consumption (6 W). Polynomial and least-squares fit algorithms are employed to remove the baseline of the spectral scan and retrieve CH4 concentrations, respectively. An Allan-Werle deviation analysis shows that the measurement precision can reach 1.4 ppb for a 60 s averaging time. Continuous measurements covering a seven-day period were performed to demonstrate the stability and robustness of the reported CH4 sensor system.

Cite

CITATION STYLE

APA

Dong, L., Li, C., Sanchez, N. P., Gluszek, A. K., Griffin, R. J., & Tittel, F. K. (2016). Compact CH4 sensor system based on a continuous-wave, low power consumption, room temperature interband cascade laser. Applied Physics Letters, 108(1). https://doi.org/10.1063/1.4939452