The influence of contour line size and location on the reproducibility of topographic measurement with the Heidelberg Retina Tomograph

8Citations
Citations of this article
15Readers
Mendeley users who have this article in their library.

Abstract

Background: The recommended contour line (CL) location with the Heidelberg Retina Tomograph (HRT) is on the inner edge of Elschnig's scleral ring. This study investigated HRT parameter reproducibility when: (i) the CL size is altered relative to Elschnig's ring; (ii) the CL is either redrawn or imported between images. Methods: using the HRT, seven 10° images were acquired for 10 normal volunteers and 10 primary open angle glaucoma (POAG) subjects. A CL was drawn on one image for each subject using Elschnig's scleral ring for reference and imported into subsequent images. The CL diameter was then (a) increased by 50μm; (b) increased by 100μm; and (c) decreased by 50μm. To investigate the effect of the method of contour line transfer between images a CL was: (1) defined for one image and imported to 6 subsequent images; (2) drawn separately for each image.Results: parameter variability improved as the size of the CL increased for the normal group relative to Elschnig's ring but was unchanged in the POAG group. The export/import function (method 1) resulted in better parameter reproducibility than the redrawing method for both groups. Conclusions: the exporting and importing function resulted in better parameter variability for both subject groups and should be used for transferring CLs across images for the same subject. Increasing the overall CL size relative to Elschnig's scleral ring improved the reproducibility of the measured parameters in the normal group. No significant difference in parameter variability was observed for the POAG group. This suggests that the reproducibility of HRT images are affected more by the variation in topography between images than change in CL definition. Copyright © 2001 The College of Optometrists.

Cite

CITATION STYLE

APA

Roff, E. J., Hosking, S. L., & Barnes, D. A. (2001). The influence of contour line size and location on the reproducibility of topographic measurement with the Heidelberg Retina Tomograph. Ophthalmic and Physiological Optics, 21(3), 173–181. https://doi.org/10.1046/j.1475-1313.2001.00561.x

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free