Jotec E-Ventus BX Stent Graft Deployment in the FEVAR and Iliac Branch Device: Single Centre Experience

  • Sayed T
  • Ahmed I
  • Rodway A
  • et al.
N/ACitations
Citations of this article
24Readers
Mendeley users who have this article in their library.

Abstract

Objectives: To evaluate the outcomes of the E-ventus BX balloon-expandable stent graft system (Jotec, Hechingen, Germany) implanted as bridging stent grafts during fenestrated endovascular aortic repair (FEVAR) and the iliac branch device (IBD) of complex aneurysms. Methods: This was a single centre retrospective analysis prospective study including all consecutive patients treated by FEVAR and the IBD performed with E-ventus BX stent grafts as bridging stents. Demographics of patients, the diameter and length of the bridging stent grafts, technical success, reinterventions, occlusions, post-operative events, and imaging (computed tomography [CT] scan and ultrasound) were prospectively collected in an electronic database. Follow-ups were performed with clinical assessment and a CT angiogram scan at four weeks after discharge followed by a duplex ultrasound every six months for two years and then a yearly duplex scan afterwards. Results: Between June 2015 and October 2017, 40 consecutive patients (three females) were treated with custom made fenestrated endografts and the iliac branch device for complex aneurysms, using the E-Ventus BX stent graft. All 82 E-Ventus BX stent grafts were successfully delivered and deployed. There was no in-hospital mortality. The early bridging stents patency rate was 97.6% (80 out of 82). The two-target vessel post-operative occlusion was secondary to kink of the renal stents and failure for re-lining of the renal artery. Of the two patients, only one needed permanent dialysis. On the late follow-up (after 30 days), two other patients demonstrated a renal stent occlusion, with one treated successfully with re-lining of the stent and the other patient treated conservatively. Neither of them needed permanent dialysis. A follow-up was maintained for 36 patients until April 2018 with a median follow-up of 18 months. All bridging stents E-Ventus BX stent grafts remained patent (78 out of 82, 95.1%). Conclusion: E-Ventus BX stent grafts used as bridging stents during FEVAR and the IBD are associated with favourable outcomes at the mid-term follow-up. Long-term follow-up is required to confirm these promising results.

Cite

CITATION STYLE

APA

Sayed, T., Ahmed, I., Rodway, A., El Sakka, K., & Yusuf, S. W. (2019). Jotec E-Ventus BX Stent Graft Deployment in the FEVAR and Iliac Branch Device: Single Centre Experience. Annals of Vascular Diseases, 12(2), 171–175. https://doi.org/10.3400/avd.oa.18-00101

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free