Abstract
Background: The impact of anesthetic equipment on clinical practice parameters associated with development of acute respiratory distress syndrome (ARDS) has not been extensively studied. We hypothesized a change in anesthesia machines would be associated with parameters associated with lower rates of ARDS. Methods: We performed a retrospective cohort study on a subset of data used to evaluate intraoperative ventilation. Patients included adults receiving a non-cardiac, non-thoracic, non-transplant, non-trauma, general anesthetic between 2/1/05, and 3/31/09 at the University of Michigan. Existing anesthesia machines (Narkomed IIb, Drager) were exchanged for new equipment (Aisys, General Electric). The initial subset compared the characteristics of patients anesthetized between 12/1/06 and 1/31/07 (pre) with those between 4/1/07 and 5/30/07 (post). An extended subset examined cases two years pre and post exchange. Using the standard predicted body weight (PBW), we calculated and compared the tidal volume (total Vt and mL/kg PBW) as well as positive end-expiratory pressure (PEEP), peak inspiratory pressure (PIP), Delta P (PIP-PEEP), and FiO2. Results: A total of 1,414 patients were included in the 2-month pre group and 1,635 patients included in the post group. Comparison of ventilation characteristics found statistically significant differences in median (pre v post): PIP (26 ± 6 v 21 ± 6 cmH2O,p < 0.001; 9.3 ± 2.2 v 9.0 ± 1.9 ml/kg predicted body weight, p
Author supplied keywords
Cite
CITATION STYLE
Blum, J. M., Davila, V., Stentz, M. J., Dechert, R., Jewell, E., & Engoren, M. (2014). Replacement of anesthesia machines improves intraoperative ventilation parameters associated with the development of acute respiratory distress syndrome. BMC Anesthesiology, 14(1). https://doi.org/10.1186/1471-2253-14-44
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.