Abstract
Most recent fine-mesh photomultipliers (FM-PMTs), with 19 dynode stages, have been tested for application to a threshold aerogel Cherenkov counter. Two properties, the gain and the pulse height resolution, have been particularly studied in a magnetic field of up to 1.5 T. The obtained results show that a FM-PMT is a suitable device for detection of aerogel Cherenkov light.
Cite
CITATION STYLE
Iijima, T., Amami, M., Adachi, I., Enomoto, R., Itoh, R., Kobayashi, S., … Yoshida, Y. (1997). Study on fine-mesh PMTs for detection of aerogel Cherenkov light. Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 387(1–2), 64–68. https://doi.org/10.1016/S0168-9002(96)00962-X