Abstract
The present work experimentally demonstrates a multimode fiber optic sensing setup for total fat detection in raw milk samples. The optical fiber arrangement incorporates a low-coherence Fabry–Perot cavity operating in dual response. The system provides a phase modulation for a total fat range from 0.97 to 4.36%. Here, the protein remains constant at ≈3%. The data indicate that maximum sensitivity close to 616 pm/%fat could be achieved at optimal wavelength operation (500 nm). In addition, the system presented a minimal repeatability error measurement of 0.08%, cross-sensitivity between protein and fat of 0.134, and a regression coefficient of r2 = 0.9763. A thermal analysis was also performed, which indicate the temperature immunity of the system. The proposed method represents a low-cost alternative to detect minimal fat variations in raw cow milk.
Author supplied keywords
Cite
CITATION STYLE
Gastélum-Barrios, A., Soto-Zarazúa, G. M., García-Trejo, J. F., Sierra-Hernandez, J. M., & Jauregui-Vazquez, D. (2019). A new method for total fat detection in raw milk based on dual low-coherence interferometer. Sensors (Switzerland), 19(20). https://doi.org/10.3390/s19204562
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.