Spectrophotometry of PRESAGE  polyurethane dosimeters

  • Krstaji N
  • Wai P
  • Adamovics J
  • et al.
N/ACitations
Citations of this article
18Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Preliminary optical density results on irradiated PRESAGE dosimeter are outlined in this article. PRESAGE is a solid dosimeter, based on a clear polyurethane combined with the leuco-dye leuco-malachite green. The purpose of these measurements was a) to obtain spectra for optimizing the wavelength of a new light source for the equipment and b) to obtain a dose-response relation. 10 PRESAGE cuvettes were given uniform doses from 0.1 to 40 Gy and later read out by spectrophotometer. The instrument used was CAMSPEC M350 Double Beam Spectrophotometer.

Cite

CITATION STYLE

APA

Krstaji, N., Wai, P., Adamovics, J., & Doran, S. (2004). Spectrophotometry of PRESAGE  polyurethane dosimeters. Journal of Physics: Conference Series, 3, 244–247. https://doi.org/10.1088/1742-6596/3/1/038