[Image quality assessment of Step&Shoot overlap reconstruction (SSOR) for cardiac computed tomography (CT)].

1Citations
Citations of this article
20Readers
Mendeley users who have this article in their library.

Abstract

Step&Shoot cardiac computed tomography (CT) provides the benefit of significant reduction in radiation dose compared to helical cardiac CT acquisitions. We think that a difference occurs in image quality by presence of overlap reconstruction (fractionated sentence, attempted to clarify). We studied the utility of Step&Shoot overlap reconstruction (SSOR). We evaluated image quality of SSOR by comparing z-axis spatial resolution on various s in scanners that SSOR is possible and for those that are impossible. SSOR indicated better z-axis spatial resolution and less variation over the field of view (FOV) (in z & x-y directions) as compared with Step&Shoot without overlap reconstruction (SS). In addition, SS showed inadequate image reproducibility due to aliasing error in z-direction (lack of sampling interval). SSOR would contribute to improvement of the image quality of Step&Shoot cardiac CT.

Cite

CITATION STYLE

APA

Chono, T., Hirano, T., Inoue, T., Tokuyasu, S., & Higuchi, K. (2011). [Image quality assessment of Step&Shoot overlap reconstruction (SSOR) for cardiac computed tomography (CT)]. Nihon Hoshasen Gijutsu Gakkai Zasshi, 67(8), 905–911. https://doi.org/10.6009/jjrt.67.905