Design and Fabrication of PDMS Microfluidic Device Combined With Urea Biosensor for Dynamic and Static Measurements

9Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

In this study, a more efficient, convenient, and cost-effective way was proposed to design and fabricate a polydimethylsiloxane (PDMS) microfluidic device using 3D printing and transfer techniques. Typically, sensor measurements can only be conducted by immersing the sensor in a solution under static conditions. However, in this study, a newly designed microfluidic device combined with a microfluidic system enables dynamic measurements with sensors. By controlling the flow and temperature of the test liquid, the physiological environment of the human body under various conditions can be simulated, resulting in more accurate and realistic data. During the dynamic condition, an average sensitivity of 6.396 mV/(mg/dL) and linearity of 0.999 were obtained at a flow rate of 75.0 μL/min. When the temperature was at 37°C, the biosensor achieved an average sensitivity of 4.381 mV/(mg/dL) and a linearity of 0.999. The dynamic response time was 7 seconds, and the drift rate was 1.47 mV/hr. The experimental results obtained under dynamic conditions showed a significant improvement in the sensing characteristics of the urea biosensor compared with the results obtained under static conditions.

Cite

CITATION STYLE

APA

Kuo, P. Y., Wang, T. H., Hsu, M. T., & Liao, C. H. (2023). Design and Fabrication of PDMS Microfluidic Device Combined With Urea Biosensor for Dynamic and Static Measurements. IEEE Journal of the Electron Devices Society, 11, 337–346. https://doi.org/10.1109/JEDS.2023.3283533

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free