Neutral Density Filters as a Tool for Cycloplegic Plusoptix-Photorefractor Measurements: An Explorative Study

  • Maria van Minderhout H
  • Joosse M
  • Schalij-Delfos N
0Citations
Citations of this article
2Readers
Mendeley users who have this article in their library.

Abstract

Purpose: The purpose of this study is to investigate the usefulness of neutral-density (ND) filters in cycloplegic-Plusoptix-photorefractor measurements. Methods: No-filter and ND-filter 0.04, 0.1 and 0.2 cycloplegic-Plusoptix-photorefractor measurements were made in 42 hypermetropic eyes. Sphere, cylinder, spherical equivalent (SEQ), J0, and J45 values were compared. Results: Mean Plusoptix-photorefractor pupil sizes were 7.7±0.68 and 7.7±0.72 mm The no-filter failure rate was 16%, with 87% in pupils >7.8 mm. Mean no-filter sphere, cylinder, SEQ, J0 and J45 values were +0.34±0.35D,-0.29±0.22D, +0.20±0.36,-0.00±0.15, and +0.02±0.11, respectively. Only ND-filter-0.04 provided 5% more successful measurements and a clinically significant alteration in the percentage of values exceeding 0.5D for sphere and SEQ (-10% and-20%), but not for cylinder (+5%). Despite the increased accuracy, 21% of the spherical outcome exceeded 0.50D. Furthermore, the single-measure-intraclass-correlation-coefficient between no-filter and ND-filter-0.04 outcome was moderate (sphere 0.78 (0.62-0.87), cylinder 0.59 (0.35-0.75), SEQ 0.68 (0.48-0.82), J0 0.73 (0.54-0.84) and J45 0.57 (0.50-0.86)) and indicated significant individual variation. Bland-Altman-analyses indicated significant bias for sphere and SEQ; p=0.038 and p=0.030. Conclusion: ND-filter-0.04 resulted in a larger proportion of successful measurements and an increased accuracy. However, an unacceptable percentage of inaccuracy was still present compared to retinoscopy. There could be validity issues with the ND-filter 0.04 or the baseline no-filter readings at the start. We conclude that cycloplegic Plusoptix-photorefraction, even with the use of a 0.04 ND filter, is not a suitable method for exact objective refraction purposes in children.

Cite

CITATION STYLE

APA

Maria van Minderhout, H., Joosse, M. V., & Schalij-Delfos, N. E. (2022). Neutral Density Filters as a Tool for Cycloplegic Plusoptix-Photorefractor Measurements: An Explorative Study. The Open Ophthalmology Journal, 16(1). https://doi.org/10.2174/18743641-v16-e2206061

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free