Abstract Introduction: The obtaining of urinary continence is due to the strength of the pelvic floor muscles (MAPs) at the moment of muscle contraction, when there are sudden increases in intra-abdominal pressure, which increases urethral closure pressure and decreases the possibility of urinary loss. Objective: To verify the reliability, type: stability, intra-examiner, of the measure of the strength of MAPs held with Peritron. Methods: Test and retest study to assess the intra-rater reliability of Peritron to measure the strength of MAPs. The sample consisted of 36 male patients, mean age 65.3 ± 7.2 years, all with urinary incontinence (UI) after radical prostatectomy. The physical therapist conducted a training for familiarization with the procedures of MAPs strength assessment with Peritron for two weeks. The strength of MAPs was measured by a perineometer of the Peritron brand (PFX 9300®, Cardio-Design Pty. Ltd, Baulkham Hills, Australia, 2153). Results: The intraclass correlation coefficient (ICC) was equal to 0.99; P = 0.0001. The typical measurement error (ETM) was equal to 3.1 cmH2O and ETM% of 4. Conclusion: Peritron showed high reliability for measuring the strength of MAPs in men, both for clinical practice and for the production of scientific knowledge. It should be noted that such measures were carried out in stability, so it is suggested that in internal consistency reliability is equivalent.Resumo Introdução: A força de contração dos músculos do assoalho pélvico (MAPs) pode contribuir para a continência urinária, quando ocorrem súbitos aumentos de pressão intra-abdominal. Objetivo: Verificar a confiabilidade, tipo estabilidade, intraexaminador, da medida da força dos MAPs, realizada com o Peritron. Métodos: Estudo de teste e reteste para avaliar a confiabilidade intra-avaliador do Peritron para medir a força dos MAPs. A amostra foi composta por 36 pacientes do sexo masculino, com média de idade de 65,3 ± 7,2 anos, todos com incontinência urinária (IU) após prostatectomia radical. O fisioterapeuta realizou um treinamento para familiarização com os procedimentos de avaliação da força dos MAPs com o Peritron, durante duas semanas. A força dos MAPs foi medida por um perineômetro da marca Peritron (PFX 9300®, Cardio-Design Pty. Ltd, Baulkham Hills, Austrália, 2153). Resultados: O CCI foi igual a 0,99; P = 0,0001. O ETM foi igual a 3,1 cmH2O e o ETM% igual a 4. Conclusão: o Peritron mostrou elevada confiabilidade para medir a força dos MAPs em homens, tanto para a prática clínica quanto para a produção de conhecimento científico. Cabe ressaltar que tais medidas foram realizadas em estabilidade, portanto, sugere-se que em consistência interna sua confiabilidade seja equivalente.Resumen Introducción: La obtención de la continencia urinaria es debido a la fuerza de los músculos del piso pélvico (MAPs) en el momento de su contracción muscular, cuando ocurren súbitos aumentos de presión intraabdominal, lo que aumenta la presión de cierre uretral y disminuye la posibilidad de pérdida de la orina. Objetivo: Verificar la confiabilidad, tipo estabilidad, intra examinador, de la medida de la fuerza de los MAPs, realizada con el Peritrón. Métodos: Estudio de prueba y reprueba para evaluar la confiabilidad intra-evaluadora del Peritrón para medir la fuerza de los MAPs. La muestra fue compuesta por 36 pacientes del sexo masculino, la media de edad era de 65,3 ± 7,2 años, todos con IU después de la prostatectomía radical. El fisioterapeuta realizó un entrenamiento para familiarización de los procedimientos de evaluación de la fuerza de los MAPs con el Peritrón, durante dos semanas. La fuerza de los MAPs fue medida por un perineómetro de la marca Peritron (PFX 9300®, Cardio-Design Pty., Ltd, Baulkham Hills, Australia, 2153). Resultados: El coeficiente de correlación intraclase (CCI) fue igual a 0,99; P = 0,0001. El error típico de la medida (ETM) fue igual a 3,1 cmH2O y el ETM% igual a 4. Conclusión: el Peritrón mostró elevada confiabilidad para medir la fuerza de los MAPs en hombres, tanto para la práctica clínica y para la producción del conocimiento científico. Cabe resaltar que tales medidas se realizaron en estabilidad, por lo que se sugiere que en consistencia interna su confiabilidad sea equivalente.
CITATION STYLE
Zaidan, P., Pereira, F. D., & Silva, E. B. da. (2018). Strength of pelvic floor in men: reliability intra examiners. Fisioterapia Em Movimento, 31(0). https://doi.org/10.1590/1980-5918.031.ao10
Mendeley helps you to discover research relevant for your work.