Performance of the MRI-based virtual bone biopsy in the distal radius: Serial reproducibility and reliability of structural and mechanical parameters in women representative of osteoporosis study populations

35Citations
Citations of this article
28Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Serial reproducibility and reliability critically determine sensitivity to detect changes in response to intervention and provide a basis for sample size estimates. Here, we evaluated the performance of the MRI-based virtual bone biopsy in terms of 26 structural and mechanical parameters in the distal radius of 20 women in the age range of 50 to 75years (mean=62.0years, S.D.=8.1years), representative of typical study populations in drug intervention trials and fracture studies. Subjects were examined three times at average intervals of 20.2days (S.D.=14.5days) by MRI at 1.5T field strength at a voxel size of 137×137×410μm 3. Methods involved prospective and retrospective 3D image registration and auto-focus motion correction. Analyses were performed from a central 5×5×5mm 3 cuboid subvolume and trabecular volume consisting of a 13mm axial slab encompassing the entire medullary cavity. Whole-volume axial stiffness and sub-regional Young's and shear moduli were computed by finite-element analysis. Whole-volume-derived aggregate mean coefficient of variation of all structural parameters was 4.4% (range 1.8% to 7.7%) and 4.0% for axial stiffness; corresponding data in the subvolume were 6.5% (range 1.6% to 13.0%) for structural, and 5.5% (range 4.6% to 6.5%) for mechanical parameters. Aggregate ICC was 0.976 (range 0.947 to 0.986) and 0.992 for whole-volume-derived structural parameters and axial stiffness, and 0.946 (range 0.752 to 0.991) and 0.974 (range 0.965 to 0.978) for subvolume-derived structural and mechanical parameters, respectively. The strongest predictors of whole-volume axial stiffness were BV/TV, junction density, skeleton density and Tb.N (R 2 0.79-0.87). The same parameters were also highly predictive of sub-regional axial modulus (R 2 0.88-0.91). The data suggest that the method is suited for longitudinal assessment of the response to therapy. The underlying technology is portable and should be compatible with all general-purpose MRI scanners, which is appealing considering the very large installed base of this modality. © 2011 Elsevier Inc.

Cite

CITATION STYLE

APA

Lam, S. C. B., Wald, M. J., Rajapakse, C. S., Liu, Y., Saha, P. K., & Wehrli, F. W. (2011). Performance of the MRI-based virtual bone biopsy in the distal radius: Serial reproducibility and reliability of structural and mechanical parameters in women representative of osteoporosis study populations. Bone, 49(4), 895–903. https://doi.org/10.1016/j.bone.2011.07.010

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free