Flow injection determination of mercury with preconcentration by amalgamation on a gold-platinum gauze by inductively coupled plasma mass spectrometry

47Citations
Citations of this article
9Readers
Mendeley users who have this article in their library.

Abstract

Trace concentrations of mercury in water samples were determined by a method involving a preconcentration procedure and inductively coupled plasma mass spectrometry (ICP-MS) detection. Mercury vapour, generated by tin(II) chloride as the reductant, was trapped by amalgamation on a gold-platinum gauze, released by controlled heating and detected by ICP-MS. A flow injection sample introduction system with time-based injection was used and the sensitivity was found to be proportional to the mass of mercury introduced. Quantitative recoveries of mercury from open ocean sea-water, coastal sea-water and fresh water reference materials were obtained. A detection limit of 200 pg l-1 for a 25 ml sample was obtained. The precision of the method was 1% (relative standard deviation) at 1 μ l-1.

Cite

CITATION STYLE

APA

Debrah, E., Denoyer, E. R., & Tyson, J. F. (1996). Flow injection determination of mercury with preconcentration by amalgamation on a gold-platinum gauze by inductively coupled plasma mass spectrometry. Journal of Analytical Atomic Spectrometry, 11(2), 127–132. https://doi.org/10.1039/JA9961100127