Abstract
Introduction: Catheter ablation is superior to drugs regarding atrial fibrillation (AF) recurrence, symptoms improvement, and mortality reduction in heart failure. POLARx™ is a novel cryoballoon, with technical improvements seeking to improve outcomes. So far, its clinical evidence is restricted to a case report. Methods: To compare the POLARx™ cryoballoon procedural safety and efficacy to the already established Arctic Front Advance PRO™ (AFAP) in a single-center cohort study, consecutive patients undergoing AF cryoablation with the POLARx™ were enrolled. Data were prospectively gathered. POLARx™ patients were compared with a historical cohort of patients submitted to AF cryoablation with the AFAP. Results: Seventy patients were analyzed, 20 in POLARx™, and 50 in the AFAP group. They all underwent first-time pulmonary vein isolation, 77% were male, 94% had paroxysmal AF, median age was 62.5 years, median CHA2DS2-VASc 1, left-atrium size 34 ml/m², and 65% were receiving anticoagulation. The primary end-point, all pulmonary veins isolation, was 100% in both groups. The complication rate was similar (0% POLARx™ vs. 5.7% AFAP, p =.39). The median total procedural time was longer in the POLARx™ group (90 min vs. 60 min, p
Author supplied keywords
Cite
CITATION STYLE
Kochi, A. N., Moltrasio, M., Tundo, F., Riva, S., Ascione, C., Dessanai, M. A., … Fassini, G. (2021). Cryoballoon atrial fibrillation ablation: Single-center safety and efficacy data using a novel cryoballoon technology compared to a historical balloon platform. Journal of Cardiovascular Electrophysiology, 32(3), 588–594. https://doi.org/10.1111/jce.14930