The anti-suck-back ability of five new high-speed air turbine handpiece models was evaluated in this study. First, suck-back pressure with water displacement within a glass tube was measured. Next, under three different conditions, how many on-off times it takes before fluorescent stains became visible on a piece of gauze at the exhaust vent was counted and the presence of fluorescent stains on the exhaust vents itself was examined. As a result, the water height for each part of one handpiece, the TWINPOWER TURBINE PAR-4HX-O, was below 0 mm. Except for under full emersion, this model, the TWINPOWER TURBINE PAR-4HX-O, did not have any visible fluorescence penetration to the exhaust vent even after 500 on-off switches under fume/mist conditions. Conversely, the other handpieces (Ti-Max X700L, T1 CONTROL, SYNEA TA-98CLED, GENTLE silence LUX 8000B) showed suck-back. In conclusion, the first mentioned new turbine handpiece, the TWINPOWER TURBINE PAR-4HX-O, had a possibility of no suck-back by itself. However, full immersion of the whole head of the handpiece which is not completely sealed must be avoided to prevent liquid intake.
Author supplied keywords
Cite
CITATION STYLE
Ozawa, T., Nakano, M., & Arai, T. (2010). In vitro study of anti-suck-back ability by themselves on new high-speed air turbine handpieces. Dental Materials Journal, 29(6), 649–654. https://doi.org/10.4012/dmj.2010-008