Inspiratory valve malfunction in a circle system: pitfalls in capnography

25Citations
Citations of this article
10Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Capnography is a useful technique in monitoring the integrity of anaesthetic equipment such as the malfunctioning of unidirectional valves in circle system. However, the lack of a precise mechanism in existing capnographs to identify the start of inspiration and the beginning of expiration in the capnograms, makes the analysis of the carbon dioxide waveforms during inspiration difficult and thus results in inaccurate assessment of rebreathing. We report a case where, during the malfunction of the inspiratory unidirectional valve in the circle system, the capnograph failed to detect the presence of substantial rebreathing. Critical analysis of the capnogram recorded during the malfunction revealed that there was substantial rebreathing which was underestimated by the capnograph as it reports only the lowest CO2 concentration rebreathed during inspiration in such abnormal situations. © 1992 Canadian Anesthesiologists.

Cite

CITATION STYLE

APA

Kumar, A. Y., Bhavani-Shankar, K., Moseley, H. S. L., & Delph, Y. (1992). Inspiratory valve malfunction in a circle system: pitfalls in capnography. Canadian Journal of Anaesthesia, 39(9), 997–999. https://doi.org/10.1007/BF03008353