Microflow Injection System for Efficient Cu(II) Detection across a Broad Range

1Citations
Citations of this article
7Readers
Mendeley users who have this article in their library.

Abstract

In this study, a modular, multi-step, photometric microflow injection analysis (micro-FIA) system for the automatic determination of Cu(II) in a bioreactor was developed. The system incorporates diverse 3D-printed modules, including a platform formed by a mixer module to mix Cu(II) with hydroxylamine, which reduces Cu(II) to Cu(I) linked to a diluter module via a Tesla valve, a chelation mixer module, a disperser module, and a detector module provided by an LED light source at λ = 455 nm and a light dependence resistor (LDR) as a light intensity detector. The system measures the color intensity resulting from the chelation between Cu(I) and neocuproine. The micro-FIA system demonstrated good capability for automatic and continuous Cu(II) determination, in a wide range of Cu concentrations, from 34 to 2000 mg L−1. The device exhibits a good repeatability (coefficient of variation below 2% across the measured concentration range), good reproducibility, and has an accuracy of around 100% between 600 and 1900 mg L−1. Real samples were analyzed using both the micro-FIA system and an atomic absorption spectroscopy method, revealing no statistically significant differences. Additionally, a Tesla valve located before the detector substituted a 3-way solenoid valve, eliminating the need for moving parts.

Cite

CITATION STYLE

APA

Ricart, D., Dorado, A. D., Lao-Luque, C., & Baeza, M. (2024). Microflow Injection System for Efficient Cu(II) Detection across a Broad Range. Chemosensors, 12(7). https://doi.org/10.3390/chemosensors12070119

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free