Abstract
AimsThe aim of the present analysis of the DATAS study was to compare the impact of dual- vs. single-chamber defibrillators on atrial fibrillation (AF) occurrence and AF-related clinical events in patients with Class I indication for implantable cardioverter defibrillators (ICDs) and no indication for dual-chamber pacing.Methods and resultsThree hundred and thirty-four patients were randomized, through a centralized assignment, to single-chamber ICDs, dual-chamber ICDs programmed as single-chamber ICDs, and dual ICDs with full diagnostics and AF prevention and therapy capabilities. The latter two groups in the first 8 months of the study, when the study design was that of a randomized parallel trial, were compared in the present analysis. The primary endpoint was composed by the following AF-related clinical events: permanent AF, AF-related hospitalizations, cardiac-embolic events, and inappropriate ICD shocks due to AF misclassification. Two hundred and twenty-three patients were available for this analysis, of whom 111 in the single-chamber-simulated group and 112 in the dual-chamber true group. Atrial fibrillation-related composite endpoint raw incidence was 9 of 111 (8.1) in the single-chamber group vs. 1 of 112 (0.9) in the dual-chamber group (P = 0.0098 by Fisher's exact test). Single-chamber ICDs were associated with a significantly higher risk to develop the AF-related composite endpoint by Cox regression analysis (hazard ratio 8.25, 95 CI 1.03-65.96, P = 0.047) and by the Kaplan-Meier survival analysis (log-rank test, P = 0.047).ConclusionDual-chamber ICDs compared with single-chamber ICDs reduced the incidence of an endpoint composed by permanent AF, AF-related hospitalizations, and ICD shocks deemed inappropriate due to AF misclassification.
Author supplied keywords
Cite
CITATION STYLE
Ricci, R. P., Quesada, A., Almendral, J., Arribas, F., Wolpert, C., Adragao, P., … Santini, M. (2009). Dual-chamber implantable cardioverter defibrillators reduce clinical adverse events related to atrial fibrillation when compared with single-chamber defibrillators: A subanalysis of the DATAS trial. Europace, 11(5), 587–593. https://doi.org/10.1093/europace/eup072
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.