Analysis of Return-to-Zero Error after the First Load of Load Cell

3Citations
Citations of this article
24Readers
Mendeley users who have this article in their library.

Abstract

The return-to-zero error of the resistance strain load cell is most obvious in the first zero-return process during loading and unloading. To improve the accuracy of the load cell, it is necessary to figure out the cause of the error. The influence of the temperature, material, and weld cup were analyzed in this paper. It was concluded that the hysteresis is the main factor affecting the return-to-zero error after the first load. The relationship between hysteresis and zero-return error after first load was obtained by a data fitting algorithm. A method to improve the return-to-zero error after the first load was proposed.

Cite

CITATION STYLE

APA

Zhuang, S., Yang, W., Cheng, X., Kevin, J. S., Liu, C., Zhang, G., … Tian, C. (2023). Analysis of Return-to-Zero Error after the First Load of Load Cell. Sensors (Basel, Switzerland), 23(21). https://doi.org/10.3390/s23218712