High throughput second harmonic imaging for label-free biological applications

  • Macias-Romero C
  • Didier M
  • Jourdain P
  • et al.
58Citations
Citations of this article
86Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Second harmonic generation (SHG) is inherently sensitive to the absence of spatial centrosymmetry, which can render it intrinsically sensitive to interfacial processes, chemical changes and electrochemical responses. Here, we seek to improve the imaging throughput of SHG microscopy by using a wide-field imaging scheme in combination with a medium-range repetition rate amplified near infrared femtosecond laser source and gated detection. The imaging throughput of this configuration is tested by measuring the optical image contrast for different image acquisition times of BaTiO₃ nanoparticles in two different wide-field setups and one commercial point-scanning configuration. We find that the second harmonic imaging throughput is improved by 2-3 orders of magnitude compared to point-scan imaging. Capitalizing on this result, we perform low fluence imaging of (parts of) living mammalian neurons in culture.

Cite

CITATION STYLE

APA

Macias-Romero, C., Didier, M. E. P., Jourdain, P., Marquet, P., Magistretti, P., Tarun, O. B., … Roke, S. (2014). High throughput second harmonic imaging for label-free biological applications. Optics Express, 22(25), 31102. https://doi.org/10.1364/oe.22.031102