Accurate measurement of cholecystokinin in plasma

294Citations
Citations of this article
70Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Shortage of reliable plasma assays has hampered studies of cholecystokinin (CCK). The assay problems are low plasma concentrations, extensive molecular heterogeneity, and close homology of CCK to gastrin, which circulates in higher concentrations. To develop an accurate CCK RIA, antibodies were raised in rabbits, guinea pigs, and mice in titers from 200 to 4 000 000. The specificity of the antisera was tested with homologous peptides, and tissue and plasma extracts. Rabbit 92128 produced antibodies in high titer (≤500 000) with sufficient avidity (K(eff)/°≤ 1012 mol-1) and the desired specificity. The antiserum binds the bioactive forms of CCK with equimolar potency and displays no reactivity with gastrin. CCK concentrations in plasma from healthy humans rose from 1.13 ± 0.10 pmol/L (mean ± SE, n = 26) to 4.92 ± 0.34 pmol/L after a mixed meal. Chromatography of human plasma revealed traces of CCK-58, a predominance of CCK-33 and CCK-22, and moderate amounts of CCK-8. The results show that it is possible to produce specific CCK-antisera using a sulfated CCK-12 analog.

Cite

CITATION STYLE

APA

Rehfeld, J. F. (1998). Accurate measurement of cholecystokinin in plasma. Clinical Chemistry, 44(5), 991–1001. https://doi.org/10.1093/clinchem/44.5.991