A comparative study is presented, analyzing quantitatively the impact of 15 shim strategies on the homogeneity of the main magnetic field over the three-dimensional breast region in 3T MRI. The results obtained in 12 female volunteers, spanning a wide range of body and breast types, indicate that the inclusion of the back and heart in the shim region of interest leads to considerable decrease in field homogeneity, and needs to be avoided. Comparison between shim strategies using volumetric B0 maps, covering the entire breast region, and 1-6 plane B0 maps indicate only minimally reduced performance for the latter. Interestingly, however, no single shim strategy relying on a limited number of B0 maps as input was found to work best in all volunteers. This was attributed to the limited capability of a small number of B0 maps to capture the B0 variability existent within breast. On the average, a rectangular shim region of interest, encompassing the breast region alone, worked best for the cohort studied here. Copyright © 2012 Wiley Periodicals, Inc.

Author supplied keywords

Cite

CITATION STYLE

APA

Hancu, I., Govenkar, A., Lenkinski, R. E., & Lee, S. K. (2013). On shimming approaches in 3T breast MRI. Magnetic Resonance in Medicine, 69(3), 862–867. https://doi.org/10.1002/mrm.24307