This article presents the design and the conception of an ultra-compact Fabry-Pérot cavity which will be used to develop an ultra-stable laser. The proposed cavity is composed of a 25 mm long ULE spacer with fused silica mirrors. It leads to an expected fractional frequency stability of 1.5 × 10-15 limited by the thermal noise. The chosen geometry leads to an acceleration relative sensitivity below 10-12 /(m/s2) for all directions.
CITATION STYLE
Didier, A., Millo, J., Lacroûte, C., Ouisse, M., Delporte, J., Giordano, V., … Kersalé, Y. (2016). Design of an ultra-compact reference ULE cavity. In Journal of Physics: Conference Series (Vol. 723). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/723/1/012029