Validity and reliability of test strips for the measurement of salivary nitrite concentration with and without the use of mouthwash in healthy adults

15Citations
Citations of this article
45Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The nitrate (NO3−)-nitrite (NO2−)-nitric oxide (NO) pathway has received considerable interest in recent years as a potential target for nutritional interventions designed to increase NO production, and elicit therapeutic effects in humans. In particular, studies have evaluated the effects of supplemental dietary NO3−, which serves as a ‘substrate’ for this pathway, on numerous different health outcomes. One challenge has been to evaluate compliance with the NO3− interventions. A recent advance in this field has been the development of a non-invasive, simple and rapid method to measure nitrite concentrations in saliva using small test salivary strips. In the present study, ten healthy adults were recruited to a randomised, crossover study and received an acute dose of NO3−-rich beetroot juice (BJ) after rinsing their mouth with either water or commercially available antibacterial mouthwash. Salivary NO3− and NO2− concentrations were measured at baseline and up to 5 h after BJ consumption using the gold-standard chemiluminescence and a colorimetric Griess assay. In addition, two salivary test strips (Berkeley Test strips, CA, USA) were used to measure NO2−concentrations at the same time points. Five observers read the strips and inter- and intra-observer reliability was measured. The Bland-Altman method was used to provide a visual representation of the agreement between the methods used to evaluate salivary NO3−/NO2−concentration. Sialin concentrations were measured at baseline and up to 5 h after BJ consumption. BJ elevated salivary NO3− and NO2− concentrations when the mouth was rinsed with water (both P < 0.01), as assessed via both chemiluminescence and Griess methods. Rinsing the mouth with antibacterial mouthwash attenuated markedly the increase in NO2− (P < 0.001), while NO3−concentrations were unaffected (P > 0.05). The Intra-Class Coefficients of Correlation (ICC) showed a high inter- and intra-observer reliability (r > 0.8). A significant positive correlation was found between absolute salivary NO2− concentrations measured by strips and Griess and chemiluminescence methods (rho = 0.83 and 0.77, respectively) and also when expressed as changes in salivary NO2− concentrations (rho = 0.80 and 0.79, respectively). Bland Altman analysis indicated a poor agreement for absolute NO2− concentrations between salivary strips and the chemiluminescence and Griess methods. A small significant negative correlation was found between changes in salivary sialin and salivary NO2ˉ concentrations (r = −0.20, P = 0.04). A non-significant positive correlation was observed between the change in salivary sialin and salivary NO3ˉ concentrations (r = 0.18, P = 0.06). This study suggests that commercially available salivary NO2− test strips provide a reasonable surrogate marker for monitoring changes in salivary NO2− concentrations in humans. However, the strips do not provide accurate estimates of absolute NO2− concentrations.

Cite

CITATION STYLE

APA

Babateen, A. M., Shannon, O. M., Mathers, J. C., & Siervo, M. (2019). Validity and reliability of test strips for the measurement of salivary nitrite concentration with and without the use of mouthwash in healthy adults. Nitric Oxide - Biology and Chemistry, 91, 15–22. https://doi.org/10.1016/j.niox.2019.07.002

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free