Abstract
In this work, a solid‐state potentiometric pH sensor is designed by incorporating a thin film of Radio Frequency Magnetron Sputtered (RFMS) Titanium Nitride (TiN) working electrode and a commercial Ag|AgCl|KCl double junction reference electrode. The sensor shows a linear pH slope of −59.1 mV/pH, R2 = 0.9997, a hysteresis as low as 1.2 mV, and drift below 3.9 mV/hr. In addition, the redox interference performance of TiN electrodes is compared with that of Iridium Oxide (IrO2) counterparts. Experimental results show −32mV potential shift (E0 value) in 1 mM ascorbic acid (reducing agent) for TiN electrodes, and this is significantly lower than the −114 mV potential shift of IrO2 electrodes with sub‐Nernstian sensitivity. These results are most encouraging and pave the way towards the development of miniaturized, cost‐effective, and robust pH sensors for difficult matrices, such as wine and fresh orange juice.
Author supplied keywords
Cite
CITATION STYLE
Shylendra, S. P., Lonsdale, W., Wajrak, M., Nur‐e‐alam, M., & Alameh, K. (2021). Titanium nitride thin film based low‐redox‐interference po-tentiometric pH sensing electrodes. Sensors (Switzerland), 21(1), 1–16. https://doi.org/10.3390/s21010042