Abstract
A method based on QuEChERS (quick, easy, cheap, effective, rugged, and safe) extraction and gas chromatography-mass spectrometry (GC-MS) detection was described for the pesticides dichlorvos, disulfoton, ethoprophos, parathion methyl, fenchlorphos, chlorpyrifos, azinphos methyl and prothiofos in sapodilla. For all compounds studies, it was observed a strong matrix effect showing the need to use matrix matched calibration strategy. Method was validated, and good linearity (R > 0.99) was obtained for all pesticides studied with limits of detection (LODs) and quantification (LOQs) ranging from 0.01 to 0.06 mg kg-1 and 0.03 to 0.2 mg kg-1, respectively. Recovery studies were performed at different levels (0.08, 0.10, 0.14, 0.20, 0.35 and 1.17 mg kg-1) and showed good results (between 70 and 120% with relative standard deviation (RSD) < 20%). A statistical test was applied to the coefficients of the analytical curves obtained in the sapodilla matrix. Analyses of commercial samples showed chlorpyrifos were detected in about 70 and 33% for fruit and pulps samples, respectively. It should be noted that chlorpyrifos is not permitted in sapodilla crops by ANVISA and EC guidelines.
Author supplied keywords
Cite
CITATION STYLE
Alcântara, D. B., Paz, M. S. O., Rodrigues, T. H. S., Fernandes, T. S. M., Barbosa, P. G. A., Loiola, A. R., … Do Nascimento, R. F. (2018). Organophosphorus pesticide in sapodilla (Manilkara Zapota) fruit. Journal of the Brazilian Chemical Society, 29(10), 2180–2188. https://doi.org/10.21577/0103-5053.20180094
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.