Abstract
We report the results of our continued review of 11 total hip arthroplasties using 22.225 mm alumina ceramic femoral heads on a Charnley flanged stem, articulating with chemically cross-linked polyethylene. There was an initial bedding-in of up to 0.41 mm at the articular surface in the first two years. This had not progressed further, at a minimum follow-up of 15 years. Radiographically no femoral or acetabular component showed loosening or osteolysis. © 2005 British Editorial Society of Bone and Joint Surgery.
Cite
CITATION STYLE
Wroblewski, B. M., Siney, P. D., & Fleming, P. A. (2005). Low-friction arthroplasty of the hip using alumina ceramic and cross-linked polyethylene. A 17-year follow-up report. Journal of Bone and Joint Surgery - Series B, 87(9), 1220–1221. https://doi.org/10.1302/0301-620X.87B9.16265