The effect of different light curing units on vickers microhardness and degree of conversion of flowable resin composites

22Citations
Citations of this article
56Readers
Mendeley users who have this article in their library.
Get full text

Abstract

This study aimed to assess the influence of different light curing units (LCUs) on the polymerization of various flowable resin composites. Three LCUs (Optilux 501, Elipar™ DeepCure-L LED and Bluephase®20i) and eight flowable resin composites: MI FIL Flow, Estelite Flow Quick, Estelite Universal Flow (medium), Estelite Universal Flow (super low), Beautifil Flow Plus, Clearfil Majesty ES Flow, Filtek Supreme Ultra flowable and TetricEvo Flow were tested. For Vickers microhardness (VHN) test and degree of conversion (DC), specimens were prepared and polymerized for 20 s. VHN test was performed at top surfaces (3 indentations) and DC for each specimen was measured using Fourier transform infrared (FT-IR) spectroscopy after 24 h dry storage in dark at 37˚C. The data were analyzed with 2-way ANOVA and t-test with Bonferroni correction. DC and hardness values showed a relationship between materials and LCUs. The curing efficacy of LCU type may depend on the material composition.

Cite

CITATION STYLE

APA

Aung, S. Z., Takagaki, T., Ikeda, M., Nozaki, K., Burrow, M. F., Abdou, A., … Tagami, J. (2021). The effect of different light curing units on vickers microhardness and degree of conversion of flowable resin composites. Dental Materials Journal, 40(1), 44–51. https://doi.org/10.4012/dmj.2019-353

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free