Abstract
Background: Mechanical ventilation may induce harmful effects in the airways of critically ill patients. Nevertheless, the effects of cyclic stretching caused by repetitive inflation-deflation of the bronchial compartment have not been well characterized in humans. The objective of the present study was to assess the effectiveness of a load-imposing device for the cyclic stretching of human bronchi. Methods Intact bronchial segments were removed from 128 thoracic surgery patients. After preparation and equilibration in an organ bath, bronchi were stretched repetitively and cyclically with a motorized transducer. The peak force imposed on the bronchi was set to 80% of each individual maximum contraction in response to acetylcholine and the minimal force corresponded to the initial basal tone before stretching. A 1-min cycle (stretching for 15 sec, relaxing for 15 sec and resting for 30 sec) was applied over a time period ranging from 5 to 60 min. The device's performance level was assessed and the properties of the stretched bronchi were compared with those of paired, non-stretched bronchi. Results Despite the intrinsic capacities of the device, the targets of the tension adjustments remained variable for minimal tension (156-178%) while the peak force set point was unchanged (87-115%). In the stretched bronchi, a time-dependent rise in basal tone (P
Cite
CITATION STYLE
Le Guen, M., Naline, E., Grassin-Delyle, S., Devillier, P., & Faisy, C. (2015). Effectiveness of a load-imposing device for cyclic stretching of isolated human Bronchi: A validation study. PLoS ONE, 10(5). https://doi.org/10.1371/journal.pone.0127765