Hematocrit compensation in electrochemical blood glucose monitoring systems

27Citations
Citations of this article
66Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Background: Hematocrit (Hct) is a common interferent in test strips used by diabetes patients to self-monitor blood glucose (BG), resulting in measurement bias. Described is an electrochemical BG monitoring system (OneTouch® Verio™) that uses a cofacial sensor design, soluble enzyme chemistry, and multiphasic waveform to efectively correct for patient Hct, delivering an accurate reading for whole BG. Methods: The test strip comprises thin-film gold and palladium electrodes arranged cofacially and spatially separated with a thin spacer. Soluble glucose-sensing reagents are located on the lower palladium electrode and are hydrated on sample application. Blood glucose is oxidized by flavoprotein glucose dehydrogenase, with electron transfer via (reduced) potassium ferrocyanide mediator at the palladium electrode. Hematocrit levels are estimated by measuring oxidation of mediator difusion to the upper gold electrode during the first portion of the assay. The Hct-corrected glucose levels are determined by an on-meter algorithm. Results: In performance testing of blood samples at five glucose levels (30-560 mg/dl) and five Hct levels (19-61%), using 12 test meters and 3 test strip lots, 100% of results (N = 2700) met International Organization for Standardization accuracy criteria (within ± 15 mg/dl and ± 20% of reference results at glucose levels of <75 and ≥75 mg/dl, respectively). Furthermore, 99.9% (2698 of 2700) of results were within ±12 mg/dl and ± 15% of reference values at glucose levels <80 and ≥80 mg/dl, respectively. Conclusions: The technology used in this system provides accurate BG measurements that are insensitive to Hct levels across the range 20-60%. © Diabetes Technology Society.

Cite

CITATION STYLE

APA

Teodorczyk, M., Cardosi, M., & Setford, S. (2012). Hematocrit compensation in electrochemical blood glucose monitoring systems. Journal of Diabetes Science and Technology, 6(3), 648–655. https://doi.org/10.1177/193229681200600320

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free