Artifact Reduction in Undersampled Projection Reconstruction MRI of the Peripheral Vessels Using Selective Excitation

14Citations
Citations of this article
18Readers
Mendeley users who have this article in their library.

Abstract

The projection reconstruction (PR)-HyperTRICKS (time resolved imaging of contrast kinetics) acquisition integrates the benefits of through-plane Cartesian slice encoding and in-plane undersampled PR. It provides high spatial resolution both inplane (about 1 mm2) and through-plane (1-2 mm), as well as relatively high temporal resolution (about 0.25 frames per second). However, undersampling artifacts that originate from anatomy superior or inferior to a coronal imaging FOV may severely degrade the image quality. In coronal MRA acquisitions, the slice coverage is limited in order to achieve high temporal resolution. In this report we describe an artifact reduction method that uses selective excitation in PR-Hyper-TRICKS. This technique significantly reduces undersampling streak artifacts while it increases the slice coverage. © 2004 Wiley-Liss, Inc.

Cite

CITATION STYLE

APA

Du, J., Thornton, F. J., Fain, S. B., Korosec, F. R., Browning, F., Grist, T. M., & Mistretta, C. A. (2004). Artifact Reduction in Undersampled Projection Reconstruction MRI of the Peripheral Vessels Using Selective Excitation. Magnetic Resonance in Medicine, 51(5), 1071–1076. https://doi.org/10.1002/mrm.20036

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free