Abstract
Inhomogeneities of the main magnetic field cause geometric distortion in echo-planar imaging, a method central to functional MRI. A number of correction methods have been proposed, most of which are based on the acquisition of a fieldmap providing the local offsets to the main magnetic field. Here, accelerated multiecho echo-planar imaging is used, with echo times short enough to enable the construction of a fieldmap of comparable quality from the data themselves. This way, each volume in a time series can be unwarped using a fieldmap obtained from that volume, avoiding volume-to-volume field-motion interactions that invalidate reference data in conventional approaches that use a single, static, fieldmap. The combination of accelerated acquisition with dynamic distortion correction yields volumes with very low distortion at repetition times similar to conventional echo-planar imaging. The method is applied to data acquired at 3 and 7 T and is shown to effectively correct image geometry. Furthermore, it is shown that dynamic distortion correction yields better temporal signal stability than correction using a static fieldmap in the presence of subject motion. Magn Reson Med 2012. © 2011 Wiley Periodicals, Inc. Copyright © 2011 Wiley Periodicals, Inc.
Author supplied keywords
Cite
CITATION STYLE
Visser, E., Poser, B. A., Barth, M., & Zwiers, M. P. (2012). Reference-free unwarping of EPI data using dynamic off-resonance correction with multiecho acquisition (DOCMA). Magnetic Resonance in Medicine, 68(4), 1247–1254. https://doi.org/10.1002/mrm.24119