Abstract

In 2011, the International Commission on Radiological Protection (ICRP) recommended a significant reduction in the lens-equivalent radiation dose limit, thus from an average of 150 to 20 mSv/year over 5 years. In recent years, the occupational dose has been rising with the increased sophistication of interventional radiology (IVR); management of IVR staff radiation doses has become more important, making real-time radiation monitoring of such staff desirable. Recently, the i3 real-time occupational exposure monitoring system (based on RaySafeTM) has replaced the conventional i2 system. Here, we compared the i2 and i3 systems in terms of sensitivity (batch uniformity), tube-voltage dependency, dose linearity, dose-rate dependency, and angle dependency. The sensitivity difference (batch uniformity) was approximately 5%, and the tube-voltage dependency was

Cite

CITATION STYLE

APA

Hattori, K., Inaba, Y., Kato, T., Fujisawa, M., Yasuno, H., Yamada, A., … Chida, K. (2023). Evaluation of a New Real-Time Dosimeter Sensor for Interventional Radiology Staff. Sensors, 23(1). https://doi.org/10.3390/s23010512