Abstract
The full-energy peak efficiency calibration and the energy resolution measurements of the 2 in . × 2 in .LaBr3γ -ray detector are presented for γ-ray energies in the 700 keV-17.6 MeV range. Measurements were done using a combination of proton-capture nuclear reactions on 27Al, 23Na, 39K, 7Li and 11B for high-energy γ-rays, and radioactive sources such as 60Co and 152Eu for the lowest energies. At high energies, two γ-rays in a cascade from proton resonance capture were employed using Al, Na2 WO4, K2 SO4 and LiBO2 targets. The obtained results were compared to the simulations performed using a GEANT4 code. © 2009 Elsevier B.V. All rights reserved.
Author supplied keywords
Cite
CITATION STYLE
Ciemała, M., Balabanski, D., Csatlós, M., Daugas, J. M., Georgiev, G., Gulyás, J., … Vitez, A. (2009). Measurements of high-energy γ-rays with LaBr3 : Ce detectors. Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 608(1), 76–79. https://doi.org/10.1016/j.nima.2009.06.019