Abstract
The present work presents an improved method to align the measurement scale mark in an immersion hydrometer calibration system of CENAM, the National Metrology Institute (NMI) of Mexico, The proposed method uses a vision system to align the scale mark of the hydrometer to the surface of the liquid where it is immersed by implementing image processing algorithms. This approach reduces the variability in the apparent mass determination during the hydrostatic weighing in the calibration process, therefore decreasing the relative uncertainty of calibration. © 2013 by the authors; licensee MDPI, Basel, Switzerland.
Author supplied keywords
Cite
CITATION STYLE
Peña-Perez, L. M., Pedraza-Ortega, J. C., Ramos-Arreguin, J. M., Tovar Arriaga, S., Aceves Fernandez, M. A., Becerra, L. O., … Vargas-Soto, J. E. (2013). Alignment of the measurement scale mark during immersion hydrometer calibration using an image processing system. Sensors (Switzerland), 13(11), 14367–14397. https://doi.org/10.3390/s131114367