Severe Underestimation of Serum Na following IVIGTreatment

1Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Current chemistry analyzers measure ion concentration using ion-selective electrodes; however, may differ in the specific technology at the bedside versus the central laboratory. Instruments utilized for point-of-care testing (POCT) at the bedside use direct ion-selective electrodes, whereas central-laboratory analyzers use indirect ion-selective electrodes. Under most circumstances, these instruments will deliver the same result; however, various substances can cause interferences in one or the other. An 18-year-old Hispanic woman with a history of immune thrombocytopenic purpura (ITP) presented at Children's National Medical Center (CNMC) with a severe headache and required intravenous immunoglobulin (IVIG) therapy. Because a discrepancy developed between her point-of-care and central-laboratory sodium values, another instrument was used to retest the central-laboratory plasma specimens. The results were more in agreement with those from the point-of-care instrument and revealed a unique interference in sodium measurement related to IVIG use.

Cite

CITATION STYLE

APA

Virk, M. S., Dean, N. P., & Wong, E. C. C. (2018). Severe Underestimation of Serum Na following IVIGTreatment. Lab Medicine, 49(4), 372–376. https://doi.org/10.1093/labmed/lmy025

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free